| ECG |
| ●Lead type:5-lead |
| ●Input: RA; LA; RL; LL; V |
| ●Sweep speed: 12.5mm/s, 25mm/s, 50mm/s |
| ●Accuracy: ±1bpm or ±1%, which is greater |
| ●Protection: Withstand 4000VAC/50Hz voltage in isolation against electrosurgical and defibrillation |
| ●S-T detection: YES, Arrhythmia analysis: YES |
| ●Alarm: YES, audible and visual alarm, alarm events recallable |
| Bitmap
|
| ●ECG, NIBP, SPO2 ●8.4″color TFT display |
| ●Suitable for adult, pediatric and neonatal patient |
| ●Basic parameters (ECG, SPO2, NIBP) in a durable case for bedside monitoring and transport monitoring |
| ●Audible and visual alarms with adjustable alarm ranges |
| ●Networkable with CONTEC central monitoring system |
| ●Powerful data management and storage capacity |
| NIBP |
| ●Method: Oscillometry |
| ●Operation modes: Manual/Automatic/STAT |
| ●Measurement unit: mmHg/KPa selectable |
| ●Measurement types: Systolic, Diastolic, Mean |
| ●Over-pressure protection: YES |
| SPO2 |
| ●Measurement range: 0-100% |
| ●Accuracy: ±2digit (70~100%); |
| ●0~69% unspecified |
| Optional: |
| ●RESP,2-TEMP,2-IBP,ETCO2,Network,Thermal Printer,Pace Maker |